Features:
- Ergonomic Design: The TOP pipette is designed with excellent ergonomics that reduce fatigue and provide a smoother and easier pipetting experience.
- Smooth and Effortless Operation: With a one-stop point of 4N and two-stop points of 10N, the TOP pipette offers a comfortable single-handed pipetting experience with smooth and effortless operation.
- Multiple Volume Range: The TOP pipette offers nine different single-channel adjustable volume models, ranging from 0.1 µl to 10 ml, making it ideal for various laboratory applications.
- Precise Volume Setting: The TOP pipette features a new piston and metal screw structure, ensuring precise suction and dispensing of liquids. Its counter increment is one thousandth of the volume range, making it easy to set and read.
- One-Click Volume Lock: The TOP pipette has a one-click volume lock feature that helps prevent pipetting errors caused by accidental volume changes, allowing for more accurate transfer of small volumes of liquids.
- Smooth Volume Adjustment: The TOP pipette offers a smooth and easy volume adjustment experience, ensuring high precision and accuracy.
- High-Quality Materials: The TOP pipette is made of a new type of material that does not require disassembly for autoclaving or UV disinfection, and is resistant to harsh chemicals.
- Individual Filter System: The TOP pipette is equipped with an individual filter system that reduces volatile reagents from entering the pipette and causing over-aspiration or corrosion of the piston.
- Easy Maintenance: The TOP pipette is easy to disassemble and comes with tools that can be used to calibrate and maintain the pipette in the laboratory.
- Online Calibration: The TOP pipette can be calibrated online, ensuring its accuracy and reliability.
Overall, the TOP pipette is a reliable, easy-to-use, and precise manual pipette that is ideal for various laboratory applications. With its ergonomic design, precise volume setting, and high-quality materials, the TOP pipette is the ideal choice for any laboratory looking for a reliable and efficient pipette.
Specification:
Volume Range | Increment | Test Volume | Systematic Error | Random Error | ||
μL | μL | μL | μL | % | μL | % |
0.1-2.5 | 0.002 | 2.5 | ±0.04 | ±1.4 | ±0.02 | ±0.7 |
1.25 | ±0.03 | ±2.5 | ±0.02 | ±1.5 | ||
0.25 | ±0.03 | ±12.0 | ±0.02 | ±6.0 | ||
0.5-10 | 0.01 | 10 | ±0.10 | ±1.0 | ±0.04 | ±0.4 |
5 | ±0.08 | ±1.5 | ±0.04 | ±0.8 | ||
1 | ±0.03 | ±2.5 | ±0.02 | ±1.8 | ||
2-20 | 0.02 | 20 | ±0.20 | ±1.0 | ±0.06 | ±0.3 |
10 | ±0.12 | ±1.2 | ±0.06 | ±0.6 | ||
2 | ±0.10 | ±5.0 | ±0.03 | ±1.5 | ||
5-50** | 0.05 | 50 | ±0.45 | ±0.9 | ±0.15 | ±0.3 |
25 | ±0.35 | ±1.4 | ±0.2 | ±0.8 | ||
5 | ±0.15 | ±3.0 | ±0.08 | ±1.6 | ||
10-100 | 0.1 | 100 | ±0.80 | ±0.8 | ±0.20 | ±0.2 |
50 | ±0.50 | ±1.0 | ±0.15 | ±0.3 | ||
10 | ±0.30 | ±3.0 | ±0.10 | ±1.0 | ||
20-200 | 0.2 | 200 | ±1.20 | ±0.6 | ±0.40 | ±0.2 |
100 | ±1.00 | ±1.0 | ±0.30 | ±0.3 | ||
20 | ±0.50 | ±2.5 | ±0.14 | ±0.7 | ||
30-300 | 0.2 | 300 | ±1.80 | ±0.6 | ±0.60 | ±0.2 |
150 | ±1.50 | ±1.0 | ±0.45 | ±0.3 | ||
30 | ±0.75 | ±2.5 | ±0.21 | ±0.7 | ||
100-1000 | 1 | 1000 | ±6.00 | ±0.6 | ±2.00 | ±0.2 |
500 | ±5.00 | ±1 | ±1.00 | ±0.2 | ||
100 | ±3.00 | ±3 | ±0.60 | ±0.6 | ||
1000-5000** | 5 | 5000 | ±30.00 | ±0.6 | ±10.00 | ±0.2 |
2500 | ±15.00 | ±0.6 | ±7.50 | ±0.3 | ||
1000 | ±12.00 | ±1.2 | ±3.00 | ±0.3 | ||
2000-10000** | 10 | 10000 | ±60.00 | ±0.6 | ±20.00 | ±0.2 |
5000 | ±40.00 | ±0.8 | ±10.00 | ±0.2 | ||
2000 | ±30.00 | ±1.5 | ±6.00 | ±0.3 |